WEBVTT

1
00:00:07.780 --> 00:00:12.240
Research film project 241, the Ventimeter.

2
00:00:17.120 --> 00:00:23.960
Fifty years ago, Dr. Francis Benedict introduced the recording spirometer, which made respiratory

3
00:00:23.960 --> 00:00:26.800
measurements practical for clinical use.

4
00:00:27.800 --> 00:00:35.160
During the early 20s, Dr. Brian Sword introduced the circle absorption system for use in anesthesia.

5
00:00:36.240 --> 00:00:42.100
These two breathing systems, the spirometer and the anesthesia circle, are similar.

6
00:00:43.020 --> 00:00:47.140
Both have separate tubes for inspiration and expiration.

7
00:00:48.420 --> 00:00:50.680
Both have unidirectional valves.

8
00:00:51.880 --> 00:00:57.120
A soda-lime canister and a reservoir to contain the respired deaths.

9
00:00:58.000 --> 00:01:05.500
In fact, the only difference between the spirometer and the circle system is in the specific reservoirs used.

10
00:01:06.620 --> 00:01:08.900
Each serves a special purpose.

11
00:01:09.980 --> 00:01:14.640
The bag does not provide a suitable means for measurement of respiration.

12
00:01:15.640 --> 00:01:21.960
The inverted bell of the spirometer does not provide a satisfactory means to breathe for the patient.

13
00:01:23.020 --> 00:01:28.920
Why not combine the advantage of these two systems and have a simple continuous measurement

14
00:01:28.920 --> 00:01:31.240
of tidal volume during anesthesia?

15
00:01:32.660 --> 00:01:40.820
We record pulse and blood pressure regularly, but respiratory rate is the only routine respiratory

16
00:01:40.820 --> 00:01:43.120
observation during anesthesia.

17
00:01:44.120 --> 00:01:50.980
To follow the patient's ventilation requires that we know tidal volume by some convenient means.

18
00:01:52.460 --> 00:01:59.440
This is a tidal volume indicator developed at Roswell Park Memorial Institute for the

19
00:01:59.440 --> 00:02:06.020
purpose of continuous measurement of the anesthetized patient's respiration.

20
00:02:06.900 --> 00:02:13.220
Whether the patient breathes spontaneously as he is now or whether we assist ventilation,

21
00:02:15.280 --> 00:02:19.680
we have continuous information as to tidal volume.

22
00:02:21.460 --> 00:02:26.680
You'll note that we have it mounted on a two chamber absorber, which incidentally we have

23
00:02:26.680 --> 00:02:31.840
found to be more efficient than our original Roswell Park absorber.

24
00:02:31.840 --> 00:02:37.960
In this circuit, the indicator has replaced the conventional pop-off valve.

25
00:02:40.400 --> 00:02:48.480
This pop-off valve supplied by the manufacturer was removed and a simple P-tube used for mounting

26
00:02:48.480 --> 00:02:49.500
the indicator.

27
00:02:50.940 --> 00:02:54.160
I'll show you the overflow valve.

28
00:02:54.160 --> 00:03:00.200
It consists of a ball and chain linked to the dome of the bellows.

29
00:03:00.200 --> 00:03:05.280
Here you can see an aluminum ball and a tapered seat.

30
00:03:07.380 --> 00:03:15.600
As the bellows is extended, the ball is lifted off its seat and allows gas to overflow to

31
00:03:15.600 --> 00:03:16.320
the outside.

32
00:03:18.320 --> 00:03:25.980
Now when we trigger the oxygen flush, note that the pressure in the circuit is not excessive.

33
00:03:27.980 --> 00:03:33.340
It seldom exceeds three or four centimeters of water.

34
00:03:35.300 --> 00:03:40.000
This then is a volume controlled pop-off valve.

35
00:03:41.140 --> 00:03:45.700
It is phased to the end expiratory period.

36
00:03:47.540 --> 00:03:52.920
The result is that with inspiration, this linkage slackens, allows the ball to fall

37
00:03:52.920 --> 00:03:55.380
into its seat and the valve becomes confident.

38
00:03:56.040 --> 00:04:02.120
So whether the patient breathes spontaneously or whether we assist inspiration, gas does

39
00:04:02.120 --> 00:04:05.080
not leave the breathing circuit during inspiration.

40
00:04:05.940 --> 00:04:10.460
Here in animation is a view to show how the ventilator works.

41
00:04:11.160 --> 00:04:17.600
As the patient breathes gas from the bellows, a corresponding volume of gas in the bag enters

42
00:04:17.600 --> 00:04:20.340
the chamber which surrounds the bellows.

43
00:04:27.000 --> 00:04:35.160
Each time the bellows moves down, this overflow valve is closed as the chain linkage slackens.

44
00:04:37.860 --> 00:04:44.180
As ventilation is assisted or controlled, pressure applied to the bag is transferred

45
00:04:44.180 --> 00:04:48.040
inside the chamber to compress the bellows.

46
00:04:48.040 --> 00:04:54.100
In other words, squeezing the bag indirectly squeezes the bellows by a volume which may

47
00:04:54.100 --> 00:04:55.680
be directly absorbed.

48
00:04:58.080 --> 00:05:04.240
Excess gas cannot be introduced into the breathing circuit because of the action of the overflow

49
00:05:04.240 --> 00:05:04.800
valve.

50
00:05:07.760 --> 00:05:15.040
For example, if the patient coughs or if the oxygen bypass is used, the overflow valve

51
00:05:15.040 --> 00:05:18.860
is lifted as soon as the bellows is fully extended.

52
00:05:21.900 --> 00:05:27.840
This action is produced by the chain linkage and occurs normally with semi-closed system

53
00:05:27.840 --> 00:05:30.400
at the end of each expiration.

54
00:05:33.300 --> 00:05:39.040
During inspiration, the breathing circuit remains closed whether a closed system or

55
00:05:39.280 --> 00:05:41.320
a high flow technique is used.

56
00:05:46.780 --> 00:05:56.620
We can use the initial inspiratory excursion as a visual guide for assisting each breath.

57
00:05:59.360 --> 00:06:08.400
For those who prefer to control the ventilation, the tidal volume indicator serves as a continuous

58
00:06:08.400 --> 00:06:13.380
monitor of the respiration by this technique.

59
00:06:15.200 --> 00:06:22.780
Now we'd like to demonstrate another advantage of the indicator, namely the basis for judging

60
00:06:22.780 --> 00:06:29.880
the adjustment of gas flow from the gas machine and the discovery of leak from the circuit,

61
00:06:30.200 --> 00:06:33.460
a very common problem in breathing equipment.

62
00:06:36.380 --> 00:06:41.160
Look at the end expiratory position of the bellows.

63
00:06:41.300 --> 00:06:46.140
See that it tends to return each time to the same position of rest.

64
00:06:47.560 --> 00:06:54.520
This indicates that we have exactly matched gas flow from the anesthesia machine with

65
00:06:54.520 --> 00:06:56.220
the patient's requirements.

66
00:06:58.340 --> 00:07:06.860
Were this flow excessive, the end position would rise to the point of overflow.

67
00:07:08.300 --> 00:07:11.420
We would increase gas flow into the circuit.

68
00:07:14.460 --> 00:07:19.460
Now we are unable to see the full tidal volume.

69
00:07:20.680 --> 00:07:29.820
Returning to a closed system with a matched flow will demonstrate the effect of leak

70
00:07:29.820 --> 00:07:34.500
by producing a leakage at the tracheal cuff.

71
00:07:37.040 --> 00:07:45.580
Although one is still able to read the patient-inspired volume, the fall of the bellows leads to

72
00:07:45.580 --> 00:07:49.440
the quick discovery that a leak has occurred.

73
00:07:54.520 --> 00:08:00.720
And as soon as it's corrected, we have that evidence as the return of the resting position

74
00:08:00.720 --> 00:08:01.460
of the bellows.

75
00:08:01.460 --> 00:08:09.300
We have found at time the indicator quite helpful in discovering the kinking of a tracheal

76
00:08:09.300 --> 00:08:09.680
tube.

77
00:08:09.980 --> 00:08:15.960
With kinkage, one will see that the volume delivered is immediately decreased.

78
00:08:18.060 --> 00:08:26.700
And with correction of the kinking, again the patient's tidal volume returns to the

79
00:08:26.700 --> 00:08:27.520
previous level.

80
00:08:29.820 --> 00:08:38.480
When we are using a closed system and with spontaneous breathing, no corrections are

81
00:08:38.480 --> 00:08:42.660
required for direct reading of the indicator scale.

82
00:08:46.720 --> 00:08:55.380
We can take the figure of 500 cc here, represented by bellows travel, as this patient's tidal

83
00:08:55.380 --> 00:08:55.660
volume.

84
00:08:56.800 --> 00:09:05.280
If we convert our circuit to our semi-closed system with a 5 liter flow rate, you will

85
00:09:05.280 --> 00:09:07.820
note that the excursions now are less.

86
00:09:08.460 --> 00:09:12.240
The bellows travel reads about 400 cc.

87
00:09:13.220 --> 00:09:18.400
Now let's discuss the correction required for the higher flow rate.

88
00:09:19.240 --> 00:09:24.100
A correction for gas entering the circuit during inspiration.

89
00:09:24.100 --> 00:09:29.480
This does not appear as a part of the travel of the bellows.

90
00:09:29.680 --> 00:09:34.340
This correction amounts to 20 cc per liter of gas flow.

91
00:09:34.820 --> 00:09:42.100
So that with a 5 liter flow, we may add 100 cc to the observed travel of the bellows.

92
00:09:47.960 --> 00:09:52.980
This would then give us a tidal volume of slightly over 500 cc.

93
00:09:52.980 --> 00:09:56.000
Now another correction is necessary.

94
00:09:56.720 --> 00:10:03.800
That due to compression of gas in the breathing circuit and the stretch of the expansible

95
00:10:03.800 --> 00:10:06.560
component, especially the breathing tubes.

96
00:10:07.740 --> 00:10:15.040
This correction amounts to about 10 cc of tidal volume per centimeter of water pressure.

97
00:10:17.240 --> 00:10:27.240
So that with a peak inspiratory pressure of 10 centimeters of water, subtract 100 cc from

98
00:10:27.240 --> 00:10:30.040
the observed travel of the bellows.

99
00:10:34.820 --> 00:10:43.200
The corrections due to gas flow from the anesthesia machine and the corrections necessary for

100
00:10:43.200 --> 00:10:48.560
pressure during controlled respiration are in the opposite direction.

101
00:10:48.620 --> 00:10:50.840
In other words, they tend to cancel each other.

102
00:10:51.300 --> 00:10:58.580
So that with a semi-closed technique and with the usual pressures required to inflate the

103
00:10:58.580 --> 00:11:05.520
anesthetized patient's lungs, we come out with a very close approximation of tidal volume

104
00:11:05.520 --> 00:11:07.040
without any correction.

105
00:11:08.040 --> 00:11:15.760
To convert from a closed system to a non-rebreathing system.

106
00:11:20.700 --> 00:11:27.500
To do this, we will occlude the inspiratory limb of the absorber,

107
00:11:32.280 --> 00:11:35.080
remove the valve Y piece,

108
00:11:41.900 --> 00:11:48.880
and we will install the double acting non-rebreathing Rubin valve.

109
00:11:51.920 --> 00:11:59.320
Now the gas flow must exactly equal the patient's minute volume for a precise adjustment.

110
00:12:01.560 --> 00:12:08.440
We can exceed this requirement of course, and again overflow will occur from the indicator.

111
00:12:12.960 --> 00:12:20.200
Now we have approximately matched the requirements with the gas flow from the machine.

112
00:12:20.200 --> 00:12:30.040
Again, we must subtract a volume due to the pressure built up in the system.

113
00:12:32.380 --> 00:12:37.140
The corrected tidal volume here would be about 650 cc.

114
00:12:38.400 --> 00:12:44.960
This device is not intended as a research instrument, but rather for clinical observation of the patient.

115
00:12:44.960 --> 00:12:52.500
The ventilometer is actually a dry spirometer mounted in the breathing circuit so as not to rob the

116
00:12:52.500 --> 00:12:58.240
anesthetist of his favorite means of breathing for the patient, namely the familiar bag.

117
00:13:02.460 --> 00:13:10.440
It incorporates an overflow valve system which replaces the conventional spring-loaded or weight-loaded pop-off.

118
00:13:11.280 --> 00:13:21.140
This design improves the accuracy of estimating the patient's tidal volume and ensures safety when a mechanical ventilator is used.

119
00:13:22.420 --> 00:13:31.920
With inspiratory pressures, gas is delivered to the patient and not through a spring-loaded or weight-loaded pop-off valve to the outside.

120
00:13:33.220 --> 00:13:39.440
All flow to the outside occurs at one time in the cycle, namely immediately after expiration.

121
00:13:40.440 --> 00:13:48.180
The ventilometer provides an objective figure of the patient's ventilation just as we record pulse and blood pressure.

